Research article| Volume 1, ISSUE 1, P86-88, August 2008

Quadruplex real-time PCR for forensic DNA quantitation

      Abstract

      Forensic DNA quantitation is an important initial step preceding PCR amplification of the STR loci even though information concerning the quality of the DNA is not revealed. A quadruplex real-time PCR (qPCR) assay was developed to quantify four DNA targets: (1) the human RB1 gene in nuclear DNA, (2) the DAZ gene present on the human Y chromosome, (3) the ATPase8 gene present in human mitochondrial DNA and (4) an artificial internal positive control to reveal possible PCR inhibition. Primers labeled with four different fluorophores are used together with a single quencher using the antiprimer quenching-based qPCR method in one reaction, in which the resultant amplicons are less than 127 bp in size. Sensitivity was shown to be less than ten copies for all four targets in the absence of amplification inhibition. The amplification remained sensitive in the presence of an excess of non-human DNA.

      Keywords

      1. Introduction

      The accurate quantitation of extracted forensic DNA is now fairly straightforward with the use of real-time PCR (qPCR). It has become an essential second step after template purification in forensic casework prior to PCR amplification of STR loci or of mitochondrial DNA (mtDNA), so as to optimize or even avoid further sample processing. There have been several reports of forensic qPCR assays [
      • Andréasson H.
      • et al.
      Real-time DNA quantification of nuclear and mitochondrial DNA in forensic analysis.
      ,
      • Richard M.L.
      • et al.
      Developmental validation of a real-time quantitative PCR assay for automated quantification of human DNA.
      ,
      • Alonso A.
      • et al.
      Real-time PCR designs to estimate nuclear and mitchondrial DNA copy number in forensic and ancient DNA studies.
      ,
      • Walker J.A.
      • et al.
      Multiplex polymerase chain reaction for simultaneous quantitation of human nuclear, mitochondrial, and male Y-chromosome DNA: application in human identification.
      ,
      • Nicklas J.A.
      • Buel E.
      • An Alu-based
      MGB Eclipse real-time PCR method for quantitation of human DNA in forensic samples.
      ,
      • Horsman K.M.
      • et al.
      Development of a human-specific real-time PCR assay for the simultaneous quantitation of total genomic and male DNA.
      ,
      • Swango K.L.
      • et al.
      A quantitative PCR assay for the assessment of DNA degradation in forensic samples.
      ] and commercial kits [
      • Green R.L.
      • et al.
      Developmental validation of the quantifiler real-time PCR kits for the quantification of human nuclear DNA samples.
      ] are now available or becoming available that enable quantification by qPCR of one or more representative targets. Most of these works describe assays which quantify up to three targets.
      An equally important issue is the presence of inhibitors that could affect subsequent processing even in the presence of adequate template concentrations. The quantitation step is an ideal moment during which the influence of such inhibitors can be estimated.
      In this work, we sought to develop a single-tube qPCR assay which allows the quantitation of three distinct but relevant forensic human DNA targets and gives information on the possible presence of PCR inhibitors. Although template DNA quantitation may not convey information concerning its quality, we endeavoured to amplify targets of less than 130 bp in size, thus giving some information as to the presence of DNA degradation.

      2. Materials and methods

      2.1 Chemistry and primers

      To reduce costs and intellectual property conflicts, we chose a novel real-time chemistry based upon an antiprimer quenching-based qPCR method [
      • Li J.
      • et al.
      Antiprimer quenching-based real-time PCR and its application to the analysis of clinical cancer samples.
      ]. The sequence of the antiprimer (AP) is TTCCCTCGGATAGCACT and is 3′-labeled with BHQ2. Primers were designed as suggested [
      • Li J.
      • et al.
      Antiprimer quenching-based real-time PCR and its application to the analysis of clinical cancer samples.
      ].

      2.1.1 Nuclear autosomal target

      The human retinoblastoma gene was selected as a single-copy target, as previously described [
      • Andréasson H.
      • et al.
      Real-time DNA quantification of nuclear and mitochondrial DNA in forensic analysis.
      ] and the primer sequences are:
      F: CCAGAAAATAAATCAGATGGTATGTAACA and R: AGTGCTATCCGAGGGAATGGTTTAGGAGGGTTGCTTCC and is 5’-labeled with FAM, such that the resultant product size is 95 bp.

      2.1.2 Y chromosomal target

      The human DAZ gene was selected of which there are normally four copies per male haploid genome [
      • Saxena R.
      • et al.
      Four DAZ genes in two clusters found in the AZFc region of the human Y chromosome.
      ], F: AGTGCTATCCGAGGGAAGAATCCTGTAACTCAGCACGTTC and is 5’-labeled with Cal Fluor Red 610 and R: TTACTATGGAAAGTGTAGATACAATTCC, such that the resultant product size is 126 bp.

      2.1.3 Mitochondrial DNA target

      The human ATPase8 gene was used as described previously [
      • Walker J.A.
      • et al.
      Multiplex polymerase chain reaction for simultaneous quantitation of human nuclear, mitochondrial, and male Y-chromosome DNA: application in human identification.
      ], F: AATATTAAACACAAACTACCACCTACCT and R: AGTGCTATCCGAGGGAATGGTTCTCAGGGTTTGTTATAATTT 5’-labeled with HEX, 96 bp amplicon.

      2.1.4 Internal positive control (IPC)

      This synthetic 77 nt oligonucleotide has the following sequence [
      • Swango K.L.
      • et al.
      A quantitative PCR assay for the assessment of DNA degradation in forensic samples.
      ]: AAGCGTGATATTGCTCTTTCGTATAGTTACCATGGCAATGCTTAGAACAATACTAATGTTGTAATCTGTCGCTATGT and is detected with these primers: F: AGTGCTATCCGAGGGAAAGCGTGATATTGCTCTTTCGTATAG which is 5’-labeled with Cy5 and R: ATAGCGACAGATTACAACATTAGTATTG.

      2.2 qPCR

      Total reaction volume was 25–40 μl. Final concentrations were 400 μM dNTPs, 5 mM Mg2+, 1 μM AP, 400 nM mtDNA primers, 200 nM remaining primers, 0.1 U/μl DNA polymerase. Thermocycles comprised 15 s at 95 °C, 30 s at 60 °C, 45 s at 50 °C during which fluorescence was measured. Multiplex signal collection was carried out using a Bio-Rad iCycler instrument.

      2.3 Standards

      Genomic DNA target standards were made by serially diluting genomic DNA previously quantified using a Nanodrop 3300 fluorospectrometer and PicoGreen. Mitochondrial DNA standards were obtained by serially diluting human DNA containing a known quantity of mitchondrial copies. IPC copy number standards were made by serially diluting a known mass of molecule previously determined by the oligonucleotide manufacturer.

      3. Results and discussion

      The above data were generated from quadruplex qPCR reactions of serially diluted human genomic and mtDNA. The reactions also contained 1.14 × 105 copies/μl of IPC. The sensitivity under these conditions is 10 copies per 40 μl qPCR reaction (Fig. 1, Fig. 2).
      Figure thumbnail gr1
      Fig. 1Amplification curves (duplicates) and serial dilution plots for RB1 and DAZ gene targets.
      Figure thumbnail gr2
      Fig. 2Amplification curves (duplicates) and serial dilution plots for ATPase8 gene, and for fixed IPC target.
      The calculated amplification efficiencies for the human DNA targets are all >100%, most likely reflecting the inherent difficulties of amplifying four targets simultaneously when present at higher copy numbers; this was not observed when carrying out singleplex amplifications of each target separately. The amplification curves observed when amplifying a target alone or as part of a multiplex were mostly superimposable. Despite extensive optimization efforts we were not able to improve upon these results and we found that the universal multiplex conditions previously described [
      • Persson K.
      • et al.
      Four-color multiplex reverse transcription polymerase chain reaction – overcoming its limitations.
      ] referring to 5′ nuclease probes did not apply in this case.
      Amplification of mtDNA under multiplex conditions inevitably gave amplification curves in which the signals decreased after reaching a maximum, as shown above, a phenomenon not observed in singleplex reactions. Agarose gel electrophoresis of the multiplex qPCR reactions did not show any additional nonspecific products. As yet we have no explanation for these findings and they could represent the ‘hook effect’ [

      Technical note no. LC 8/99, Roche Molecular Biochemicals, 1999.

      ] which has no effect on the quantitation.
      Amplification of the fixed IPC under multiplex conditions varied by a maximum of one Ct regardless of the varying copy numbers of the other three targets, as shown above. This shows that, under these conditions, IPC signal generation is independent of that of the other three targets and that the IPC is functioning as anticipated. Preliminary experiments (data not shown) demonstrated that when the IPC is spiked into samples known to have PCR inhibitors, its amplification profile reflects this fact in terms of the shape of the amplification curve and in terms of the increased Ct.
      Future work will be directed to cloning the three genomic targets so as to permit setting up the standard curves more easily and reliably using plasmids.

      Conflict of interest

      None.

      References

        • Andréasson H.
        • et al.
        Real-time DNA quantification of nuclear and mitochondrial DNA in forensic analysis.
        BioTechniques. 2002; 33: 402-411
        • Richard M.L.
        • et al.
        Developmental validation of a real-time quantitative PCR assay for automated quantification of human DNA.
        J. Forensic Sci. 2003; 48: 1041-1046
        • Alonso A.
        • et al.
        Real-time PCR designs to estimate nuclear and mitchondrial DNA copy number in forensic and ancient DNA studies.
        Forensic Sci. Int. 2004; 139: 141-149
        • Walker J.A.
        • et al.
        Multiplex polymerase chain reaction for simultaneous quantitation of human nuclear, mitochondrial, and male Y-chromosome DNA: application in human identification.
        Anal. Biochem. 2005; 337: 89-97
        • Nicklas J.A.
        • Buel E.
        • An Alu-based
        MGB Eclipse real-time PCR method for quantitation of human DNA in forensic samples.
        J. Forensic Sci. 2005; 50: 1081-1090
        • Horsman K.M.
        • et al.
        Development of a human-specific real-time PCR assay for the simultaneous quantitation of total genomic and male DNA.
        J. Forensic Sci. 2006; 51: 758-765
        • Swango K.L.
        • et al.
        A quantitative PCR assay for the assessment of DNA degradation in forensic samples.
        Forensic Sci. Int. 2006; 158: 14-26
        • Green R.L.
        • et al.
        Developmental validation of the quantifiler real-time PCR kits for the quantification of human nuclear DNA samples.
        J. Forensic Sci. 2005; 50: 809-825
        • Li J.
        • et al.
        Antiprimer quenching-based real-time PCR and its application to the analysis of clinical cancer samples.
        Clin. Chem. 2006; 52: 624-633
        • Saxena R.
        • et al.
        Four DAZ genes in two clusters found in the AZFc region of the human Y chromosome.
        Genomics. 2000; 67: 256-267
        • Persson K.
        • et al.
        Four-color multiplex reverse transcription polymerase chain reaction – overcoming its limitations.
        Anal. Biochem. 2005; 344: 33-42
      1. Technical note no. LC 8/99, Roche Molecular Biochemicals, 1999.