Research article| Volume 1, ISSUE 1, P80-82, August 2008

Development and usage of a NIST standard reference material for real time PCR quantitation of human DNA

      Abstract

      National Institute of Standards and Technology SRM 2372 human DNA quantitation standard has been produced to support the need for a human-specific DNA quantitation standard in forensic casework and calibration of new quantitative polymerase chain reaction (qPCR) assays. The conventional DNA concentration has been assigned with one of the U.S. National Reference UV/Visible Spectrophotometers, assuming an absorbance of 1.0 at 260 nm equals 50 ng/μL of double stranded DNA. In addition, an interlaboratory study has been conducted, to verify that the SRM 2372 materials perform well in currently used DNA quantitation assays by the forensic DNA community. Each unit of SRM 2372 consists of three well-characterized DNA extracts. Component A is a single-source human male material derived from blood. Component B is a multiple-source human female material derived from blood. Component C was purchased as a purified unsheared genomic human DNA (Sigma–Aldrich Co., St. Louis, MO) obtained as a lyophilized human genomic extract and has both male and female donors. SRM 2372 is intended to enable the comparison of DNA concentration measurements across time and place. Manufacturers can use SRM 2372 to validate the values assigned to their own reference materials. Individual forensic laboratories can use SRM 2372 to validate DNA quantitation methods and to verify the assigned concentration of in-house or commercial DNA calibration standards.

      Keywords

      1. Introduction

      Forensic DNA identification laboratories rely on reproducible quantitation of extracted human DNA for downstream analysis. Quantitation inaccuracy leads to too much or too little sample being used in the polymerase chain reaction (PCR) amplification process that may lead to artefacts, wasted time, resources, and potential loss of critical sample.
      NIST has conducted several interlaboratory studies to assess the “state of the art” in human DNA quantitation methods used by the forensic DNA community and to determine whether or not a standard reference material for DNA quantification would be useful. In NIST Quantitation Study 2004 (QS04), 80 participants provided 287 independent sets of results from 19 different quantification methods [
      • Kline M.C.
      • Duewer D.L.
      • Redman J.W.
      • Butler J.M.
      Results from the NIST 2004 DNA Quantitation Study.
      ]. The expected one standard deviation among-laboratory variability for sub-ng/μL DNA concentration samples in QS04 is about a factor of 1.8. (A factor-of-two uncertainty in the amount of template DNA is equivalent to ± a single PCR amplification cycle.) Much of this variability apparently resulted from systematic biases among the 10 quantitative PCR (qPCR) methods used in QS04. Members of the forensic community have also communicated problems with lot-to-lot variability of commercial quantitation calibrators included in commercial human DNA quantitation kits.
      The weight of the evidence from QS04 suggests that use of a human DNA standard should help reduce among-laboratory, among-calibrant, and among-method quantitation variability. SRM 2372 human DNA quantitation standard was designed to meet this need.

      2. Materials and methods

      2.1 Material preparation

      SRM 2372 contains three components labelled A, B, and C. All components are human genomic DNA. Component A was prepared at NIST from Buffy coat white blood cells from single-source anonymous male. Component B was prepared at NIST from Buffy coat cells from multiple anonymous female donors. Component C was obtained as a commercial lyophilized human genomic extract and has both male and female donors.
      The method used to isolate genomic DNA from the white blood cells found in the human Buffy coats is based on a modified “salting out” procedure [
      • Miller S.A.
      • Dykes D.D.
      • Polesky H.F.
      A simple salting out procedure for extracting DNA from human nucleated cells.
      ]. After the initial extraction, the material was re-extracted to assure purity. In addition, Component A was treated with RNase before re-extraction [
      • Sambrook J.
      • Russell D.W.
      Molecular Cloning a Laboratory Manual.
      ].

      2.2 Decadic attenuance measurements

      The nominal concentration of a DNA solution is derived from the widely accepted assertion that for a solution of double stranded DNA, an optical density (OD) at 260 nm of 1.0 corresponds to a DNA concentration of 50 μg/mL (50 ng/μL) [
      • Sambrook J.
      • Russell D.W.
      Molecular Cloning a Laboratory Manual.
      ,
      • Stulnig T.M.
      • Amberger A.
      Exposing contaminating phenol in nucleic acid preparations.
      ]. Optical densities at four additional wavelengths (230 nm, 270 nm, 280 nm and 330 nm) are also traditionally used in the assessment of DNA quality [
      • Sambrook J.
      • Russell D.W.
      Molecular Cloning a Laboratory Manual.
      ]. The SRM 2372 component materials are therefore certified for Decadic Attenuance at 230 nm, 260 nm, 270 nm, 280 nm and 330 nm using UV/vis spectrophotometry. Measurements were performed on the NIST high accuracy reference spectrophotometer. Values at 260 nm along with nominal DNA concentrations for each SRM component are listed in Table 1.
      Table 1Decadic attenuance and nominal DNA concentration values for the SRM 2372 (n = 3)
      Component260 nmError at 260 nmNominal [DNA] (ng/μL)
      A1.049±0.02552.5
      B1.073±0.03053.6
      C1.086±0.02854.3

      3. Results and discussion

      3.1 Interlaboratory testing

      A total of 32 laboratories participated in the interlaboratory testing of the materials. Participants ran their qPCR method of choice using the SRM materials. SRM component C was used as the DNA standard to determine the concentration of components A and B. The following eight DNA quantification methods were used in the study: Quantifiler [

      Quantifiler Human DNA Quantitation Kit and Quantifiler Y Human Male DNA Quantitation Kit User's Manual, Part Number 4344790, Rev. A., Applied Biosystems, Foster City, CA, 2003.

      ], Centre of Forensic Sciences [
      • Richard M.L.
      • Frappier R.H.
      • Newman J.C.
      Developmental validation of a real-time quantitative PCR assay for automated quantification of human DNA.
      ], Sybr green ALU [
      • Nicklas J.A.
      • Buel E.
      Development of an Alu-based, real-time PCR method for quantitation of human DNA in forensic samples.
      ], Probe-based ALU [
      • Nicklas J.A.
      • Buel E.
      Simultaneous determination of total human and male DNA using a duplex real-time PCR assay.
      ], Picogreen [
      • Singer V.L.
      • Jones L.J.
      • Yue S.T.
      • Haugland R.P.
      Characterization of PicoGreen reagent and development of a fluorescence-based solution assay for double-stranded DNA quantitation.
      ], CA-DOJmonoTH01 [
      • Timken M.D.
      • Swango K.L.
      • Orrego C.
      • Buoncristiani M.R.
      A duplex real-time qPCR assay for the quantification of human nuclear and mitochondrial DNA in forensic samples: implications for quantifying DNA in degraded samples.
      ], CA-DOJ-triplexTH01 [
      • Timken M.D.
      • Swango K.L.
      • Orrego C.
      • Buoncristiani M.R.
      A duplex real-time qPCR assay for the quantification of human nuclear and mitochondrial DNA in forensic samples: implications for quantifying DNA in degraded samples.
      ], CA-DOJ-triplexCSF [
      • Swango K.
      • Timken L.
      • Chong M.D.
      • Buoncristiani M.R.
      A quantitative PCR assay for the assessment of DNA degradation in forensic samples.
      ]. The results for the majority of the methods fall within a 95% confidence level (data not shown). The net result of the study is that the SRM materials are appropriate for use with different qPCR methods.

      3.2 Value assignment for DNA standard calibration

      An example of DNA standard value assignment is shown in Table 2. DNA concentration values are listed as ng/μL. Four commercial human DNA standards were calibrated using the SRM materials. Samples were evaluated using the Quantifiler qPCR method. Experiments were carried out in duplicate over dilutions ranging from 10× to 200× of the stock materials. Commercial standards 1–3 were found to deviate greater than 30% from the stated concentration supplied with the material.
      Table 2DNA standard calibration using SRM 2372
      Commercial standard
      1234
      [DNA]S.D.[DNA]S.D.[DNA]S.D.[DNA]S.D.
      Dilution
       10×1053.212211265.825610.1
       50×1053.31227.31450.82727.8
       100×996.211311.61380.527010.5
       200×1001.713718.51373.93113.7
       Assigned102123136277
       Stated200200200260
       Deviation (%)−49−38−326

      3.3 Summary

      The three SRM components are highly purified human DNA extracts certified for decadic attenuance (absorption), but a nominal DNA concentration can easily be obtained assuming 1 OD = 50 ng/μL of double stranded DNA. The materials are intended to be used as calibrants for standards used in quantitative real time PCR assays. Commercial or in-house DNA standards can be calibrated using the material. The material is not intended for everyday use as a standard. The materials are to be stored at 4 °C and not frozen. SRM 2372 provides an independent means to calibrate the concentration of a DNA standard. DNA standards obtained from different vendors or varying lots can be monitored for concentration accuracy and potential performance variation. Further details about SRM 2372 can be found at http://www.cstl.nist.gov/biotech/strbase/srm2372.htm.

      Funding Source

      National Institute of Justice through interagency agreement 2003-IJ-R-029 with the NIST Office of Law Enforcement Standards. The sponsor had no involvement in the development of the paper or decisions related to the paper.

      Conflict of interest

      None.

      References

        • Kline M.C.
        • Duewer D.L.
        • Redman J.W.
        • Butler J.M.
        Results from the NIST 2004 DNA Quantitation Study.
        J. Forensic Sci. 2005; 50: 571-578
        • Miller S.A.
        • Dykes D.D.
        • Polesky H.F.
        A simple salting out procedure for extracting DNA from human nucleated cells.
        Nucleic Acids Res. 1988; 16: 1215
        • Sambrook J.
        • Russell D.W.
        Molecular Cloning a Laboratory Manual.
        Cold Spring Harbor Laboratory Press, Cold Spring Harbor, New York2001
        • Stulnig T.M.
        • Amberger A.
        Exposing contaminating phenol in nucleic acid preparations.
        Biotechniques. 1994; 16: 403-404
      1. Quantifiler Human DNA Quantitation Kit and Quantifiler Y Human Male DNA Quantitation Kit User's Manual, Part Number 4344790, Rev. A., Applied Biosystems, Foster City, CA, 2003.

        • Richard M.L.
        • Frappier R.H.
        • Newman J.C.
        Developmental validation of a real-time quantitative PCR assay for automated quantification of human DNA.
        J. Forensic Sci. 2003; 48: 1041-1046
        • Nicklas J.A.
        • Buel E.
        Development of an Alu-based, real-time PCR method for quantitation of human DNA in forensic samples.
        J. Forensic Sci. 2003; 48: 936-944
        • Nicklas J.A.
        • Buel E.
        Simultaneous determination of total human and male DNA using a duplex real-time PCR assay.
        J. Forensic Sci. 2006; 51: 1005-1015
        • Singer V.L.
        • Jones L.J.
        • Yue S.T.
        • Haugland R.P.
        Characterization of PicoGreen reagent and development of a fluorescence-based solution assay for double-stranded DNA quantitation.
        Anal. Biochem. 1997; 249: 228-238
        • Timken M.D.
        • Swango K.L.
        • Orrego C.
        • Buoncristiani M.R.
        A duplex real-time qPCR assay for the quantification of human nuclear and mitochondrial DNA in forensic samples: implications for quantifying DNA in degraded samples.
        J. Forensic Sci. 2005; 50: 1044-1060
        • Swango K.
        • Timken L.
        • Chong M.D.
        • Buoncristiani M.R.
        A quantitative PCR assay for the assessment of DNA degradation in forensic samples.
        Forensic Sci. Int. 2006; 158: 14-26