Forensic Science International: Genetics Supplement Series
Volume 2, Issue 1 , Pages 71-73, December 2009

Automated washing of FTA Card punches and PCR setup for reference samples using a LIMS-controlled Sias Xantus automated liquid handler

Section of Forensic Genetics, Department of Forensic Medicine, Faculty of Health Sciences, University of Copenhagen, DK-2100 Copenhagen, Denmark

Received 3 August 2009; accepted 3 August 2009. published online 01 September 2009.

Abstract 

We have implemented and validated automated methods for washing FTA Card punches containing buccal samples and subsequent PCR setup using a Sias Xantus automated liquid handler. The automated methods were controlled by worklists generated by our LabWare Laboratory Information Management System (LIMS). The protocols were validated according to EN/ISO 17025 for use with STR amplifications kits AmpFSTR® Identifiler® and Y-filer® (Applied Biosystems).

Keywords: Laboratory automation, LIMS, FTA, Reference samples

 

PII: S1875-1768(09)00015-8

doi:10.1016/j.fsigss.2009.08.002

Forensic Science International: Genetics Supplement Series
Volume 2, Issue 1 , Pages 71-73, December 2009